Rohde & Schwarz ESR7B3-B4-B5-B9-B18-B22-B29-B50-K53 EMI Test Receiver, 9 kHz to 7 GHz
Manufacturer Rohde & SchwarzDescription
The R&S ESR7 EMI Test Receiver is a dual instrument that measures electromagnetic disturbances while simultaneously functioning as a fully featured, powerful signal and spectrum analyzer for lab applications. Rapid-fire-speed measuring occurs by the instrument’s use of an FFT-based time domain scan and detailed analysis and history of disturbance signals is accomplished by realtime spectrum analysis using a wide variety of diagnostic tools. This robust and compact, dual instrument is user friendly with its straightforward display, intuitive touchscreen interface.
Features and Specifications:
Features
-
- EMI test receiver and signal/spectrum analyzer combined in a single instrument
- Compliant with CISPR 16-1-1 Ed. 3.1
- Preselection with integrated 20 dB preamplifier
- Ultra-fast time domain scan in addition to conventional stepped frequency scan
- Realtime spectrum analysis with up to 40 MHz span for detailed investigation of disturbance signals
- Powerful measurement and analysis functions in frequency and time domain
- Time domain display with high resolution (50 μs)
- Automatic test routines
- Convenient operation, straightforward display
- Robust and compact – ideal for the mobile use
Specifications
-
- Frequency
- Frequency Range: 9 kHz to 7 GHz
- Level
- Max RF level (CW):
- RF attenuation ≥ 10 dB; RF preamplifier off: 30 dBm (= 1 W)
- RF attenuation ≥ 10 dB; RF preamplifier on: 23 dBm (= 0.2 W)
- Max Pulse Voltage:
- RF attenuation ≥ 10 dB:
- Input 1: 150 V
- Input 2: 450 V
- RF attenuation ≥ 10 dB:
- Max Pulse Energy:
- RF attenuation ≥ 10 dB; 10 μs
- Input 1: 1 mWs
- Input 2: 20 mWs
- RF attenuation ≥ 10 dB; 10 μs
- 1 dB Compression:
- RF attenuation 0 dB; RF preamplifier and preselection off: +3 dBm, nominal
- Max RF level (CW):
- IF and Resolution Bandwiths
- Analyzer Mode (span ≥ 10 Hz) and Receiver Mode: 10 Hz to10 MHz (–3 dB) in 1/2/3/5/10 steps
- Analyzer and Receiver Mode: 200 Hz, 9 kHz, 120 kHz (–6 dB), 1 MHz (impulse bandwidth)
- Preselection:
- Can be switched off in analyzer mode: 16 Fixed Filters
- Preamplifier:
- Can be switched on/off: 1 kHz to 7 GHz, 20 dB gain, nominal
- Measurement Time:
- Analyzer Mode (Sweep Time):
- span = 0 Hz: 1 μs to 16 000 s
- span ≥ 10 Hz (swept): 1 ms to 16 000 s
- span ≥ 10 Hz (FFT): 7 μs to 16 000 s
- Receiver Mode (Stepped Frequency Scan): 50 μs to 100 s (per frequency)
- Receiver Mode (Time Domain Scan): 50 μs to 100 s (per frequency subrange)
- Analyzer Mode (Sweep Time):
- Frequency Step Size:
- Receiver Mode (Stepped Frequency Scan): min. 1 Hz
- Receiver Mode (Time Domain Scan): 0.25 × IF bandwidth
- Detectors
- Receiver Mode:
- maximum peak
- minimum peak
- quasi-peak
- RMS
- average
- average with meter time constant (CISPR average)
- RMS average (CISPR-RMS)
- Receiver Mode:
- Displayed Average Noise Level (DANL)
- Receiver Mode, Nominal, Average Detector (AV), RF Attenuation 0 dB, Termination 50 Ω:
- Preamplifier Off:
- 500 MHz, bandwidth 120 kHz: < 6 dBμV
- 3 GHz, bandwidth 1 MHz: < 17 dBμV
- Preamplifier On:
- 500 MHz, bandwidth 120 kHz: < –7 dBμV
- 3 GHz, bandwidth 1 MHz: < 5 dBμV
- Preamplifier Off:
- Receiver Mode, Nominal, Average Detector (AV), RF Attenuation 0 dB, Termination 50 Ω:
- Number of Sweep (Trace) Points
- Analyzer Mode (Standard): 101 to 32 001
- Analyzer Mode (EMI): 101 to 200 001
- Receiver Mode: Max 4 000 000
- Realtime Analyzer (Option): 801
- Total Measurement Uncertainty
- CW signal, level 0 dB to –70 dB below reference level, S/N > 20 dB, auto sweep time, RF attenuation 10 dB, 20 dB, 30 dB, 40 dB, preselection on,
span/RBW<100, 95% confidence level, +200C to +300C- 9 kHz ≤ f < 3.6 GHz: 0.46 dB
- 3.6 GHz ≤ f ≤ 7 GHz: 0.57 dB
- CW signal, level 0 dB to –70 dB below reference level, S/N > 20 dB, auto sweep time, RF attenuation 10 dB, 20 dB, 30 dB, 40 dB, preselection on,
- Frequency
Options (included)
| R&S®ESR-B1 | Ruggedized Housing (corner guards and front panel cover) for R&S®ESR |
| R&S®FSV-B4 | OCXO, Precision Reference Frequency |
| R&S®FSV-B9 | Tracking Generator 100 kHz to 4 GHz/7 GHz |
| R&S®ESR-B18 | Solid State Drive (SSD, removable) |
| R&S®FSV-B22 | Preamplifier, 9 kHz to 4 GHz/7 GHz |
| R&S®ESR-B29 | Frequency Extension 10 Hz, including EMI bandwidths in decade steps |
| R&S®ESR-B50 | Hardware for Time Domain Scan and Realtime Analysis |
| R&S®ESR-K53 | Time Domain Scan (requires R&S®ESR-B50) |
Calibration
Standard calibration with a 90-day warranty included. Please contact us for additional calibration service prices.
NIST-Traceable Calibrations
Standard Calibration Price (without data): Request a Quote
ANSI Z540 Calibration Price (with as-found & as-left data): Request a Quote
ISO/IEC 17025:2005 Accredited Calibration Price (with data & uncertainties): Request a Quote
ANSI Z540.3 with data & uncertainties and Guardbanding: Request a Quote
Test & Measurement Equipment Repair Services: Request a Quote